Preaparation of Ammonium standard Solution (1ppm) For Limit Test

PROCEDURE
Dilute 10.0 ml of a 0.0741% w/v solution of ammonium chloride to 25.0 ml with ammonia-free     water. Dilute 1 Volume of the resulting solution to 100 volumes with ammonia-free water, immediately before use.
                                                            